X-ray gas-solid phase high-temperature and high-pressure in-situ device
Device parameters;
1.The device can realize the synchrotron radiation X-ray transmission test mode
2. The device includes a temperature controller, an in-situ pool cavity, a heating system, and upper computer software
3. The device adopts PID intelligent control, USB signal transmission, temperature curve data collection, which can be stored and retrieved.
4. Multi-point temperature control correction is available, temperature can be set within limits, and temperature can be set in multiple program segments.
5. Temperature range: Room temperature -600℃, temperature control accuracy ±0.1℃, heating rate: > 100℃/min
6. The device adopts resistance heating, with the sensor placed beneath the sample, which can directly reflect the true temperature changes of the sample
9. The device adopts O-ring sealing, which has a good seal, can be evacuated and filled with reaction gas
10. The device is made of aviation aluminum, with an external dimension of 120mm*70mm*35mm and a net weight of less than 1Kg.
11.Cooling method of 11 chambers: Circulating water cooling is adopted;