Synchrotron‑Radiation X‑Ray Gas‑Liquid High‑Temperature In‑situ Device

SAXS Gas‑Liquid High‑Temperature In‑situ Device
Device Parameters (Temperature customizable):
1. The device is applicable to in‑situ small‑angle X‑ray scattering (SAXS) tests with synchrotron radiation.
2. The device consists of temperature controller, in‑situ cell chamber, heating system and upper‑computer software.
3. PID intelligent control and USB signal transmission are adopted. Temperature curve data can be collected, stored and retrieved.
4. Multi‑point temperature calibration, temperature limit setting and multi‑segment temperature programming are supported.
5. Temperature range: room temperature‑200℃; temperature‑control accuracy: ±1℃; heating rate: >50℃/min.
6. Resistance heating is adopted. The sensor is immersed inside the liquid to accurately reflect the real‑time temperature change of the sample.
7. Liquid volume: about 15 mL; window diameter: 6 mm.
8. Sealed with O‑ring for good sealing performance, and reaction gas can be introduced.
9. Made of aviation aluminum. Overall dimension: 120 mm*70 mm*35 mm; net weight: less than 1 kg.
10. Chamber cooling mode: circulating‑water cooling.
11. Easy‑to‑operate and convenient for installation.